Vue de la page 70
57
Bird 5012B, 5016B, 5017B, 5018B, and 5019B Wideband Power
Sensors
Sensor Characteristics
Frequency Range
5012B
5016B
5017B
5018B
5019B
350 MHz - 4.0 GHz
350 MHz - 4.0 GHz
25 MHz - 1.0 GHz
150 MHz - 4.0 GHz
25 MHz -1.0 GHz
RF Power Range
5012B
5016B
5017B
5018B
5019B
150 mW - 150 W Avg. 400 Watts Peak
25 mW - 25 W Avg. 60 Watts Peak
500 mW - 500 W Avg. 1300 Watts Peak
100 mW - 25 W Avg. 60 Watts Peak
100 mW - 100 W Avg. 260 Watts Peak
Maximum Power See Figure 7 on page 60
Impedance, Nominal 50 ohms
Insertion Loss, Max
0.35 – 1 GHz
1 – 4 GHz
0.05 dB
0.1 dB
Input VSWR, Max
0.35 – 2.5 GHz
2.5 – 4 GHz
1.05:1
1.10:1
Directivity, Min
5012B
5016B
5017B
5018B
5019B
30 dB up to 3 GHz,
28 dB from 3 to 4 GHz
30 dB up to 3 GHz,
28 dB from 3 to 4 GHz
28 dB up to 100 MHz,
30 dB from 100 to 1000 MHz
30 dB up to 3 GHz,
28 dB from 3 to 4 GHz
28 dB up to 100 MHz,
30 dB from 100 to 1000 MHz
RF Connectors N Female
Interface:
DPM
PC Serial Port
PC USB Port
Male DB-9, EIA-232, 9600 baud, no
parity, 8 data bits, 1 stop bit
Fem. DB-9, EIA-232, 9600 baud, no
parity, 8 data bits, 1 stop bit
USB 1.1 interface
Power Supply:
DPM
USB Port
DC Connector
From host instrument via cable
less than one low-power USB load
7 – 18 VDC, < 100 mA