PUBLIC SAFETY
SITE ANALYZER™
CABLE AND ANTENNA TESTER
FOR WIRELESS SYSTEMS
OPERATING INSTRUCTIONS
©Copyright 2002 by Bird Electronic Corporation
Instruction Book Part Number 920-7002A170S Rev. A
Site Analyzer is a trademark of Bird Electronic Corporation
Microsoft® and Windows® are registered trademarks
of the Microsoft Corporation
Vue de la page 0
1 2 3 4 5 6 ... 113 114

Résumé du contenu

Page 1 - SITE ANALYZER™

PUBLIC SAFETYSITE ANALYZER™CABLE AND ANTENNA TESTERFOR WIRELESS SYSTEMSOPERATING INSTRUCTIONS©Copyright 2002 by Bird Electronic CorporationInstruction

Page 2 - This page is not blank

viiiFeatures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87Computer Requirements . . . . . . . . . . .

Page 3 - Safety Precautions

Bird Site Analyzer90TroubleshootingAny service procedure not covered in this manual should be referred to an authorized service facility. Locate the p

Page 4

Maintenance91Keys do not respond Unit is “Locked Up” Turn the unit off and then back on.Perform a full system initialization. Disconnect the AC adapte

Page 5

Bird Site Analyzer92Battery Replacement! Lay the Site Analyzer, display side down, on a clean surface.! Remove the four screws and the battery cover.

Page 6 - About This Manual

Maintenance93Unit Reset! Lay the Site Analyzer, display side down, on a clean surface.! Remove the battery. Follow the instructions in “Battery Replac

Page 7

Bird Site Analyzer94Flash ROM UpgradeThe Site Analyzer uses flash-programmable ROM. It is not necessary to replace the EPROM to upgrade the firmware.☞

Page 8 - Table of Contents

Maintenance95Customer ServiceIf you need to return the unit for any reason, contact the Bird Service Center for a return authorization. All instrument

Page 9

Bird Site Analyzer96SpecificationsAntenna TestingFrequency CharacteristicsFrequency Range: 25 MHz to 1700 MHzFrequency Resolution:25 – 800 MHz800 – 17

Page 10

Maintenance97Maximum Input Signal(Damage Level)≥ 22 dBmOutput Power < 0 dBmDistance to Fault (DTF) MeasurementMode: InternalResolution [R] X-Axis:W

Page 11 - Chapter 1 Introduction

Bird Site Analyzer98Power Measure Mode (SA-1700P Only)Function: Displays power from Bird power sensors, VSWR alarm and BPM (specifications determined

Page 12 - Site Analyzer Features

Maintenance99International StandardsEMC: Conforms to: EN 63126-1:1997Safety: Conforms to: EN 61010-1:1993, withAmendment A2:1995CalibrationCalibration

Page 13 - Introduction

1Chapter 1 IntroductionThe Bird Site Analyzer is a multifunction test instrument for use in installation and maintenance of wireless systems.Antenna s

Page 14 - Component Description

Bird Site Analyzer100Measurement Uncertainty (Return Loss after one refresh)

Page 15

Maintenance101Optional Equipment AvailablePrecision Open/Short/Load Combination:Connector Type Part NumberMale NFemale NMale 7/16 DINFemale 7/16 DINCA

Page 16

Bird Site Analyzer102Hard TransitCaseP/N: 7002C870Holds Site Analyzer, standard accessories, calibration combo, armored cable assembly, and adapters.E

Page 17 - Connection Description

Limited WarrantyAll products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of two (2) years, u

Page 18 - Display Description

DECLARATION OF CONFORMITYManufacturer: Bird Electronic Corporation30303 Aurora RoadCleveland, Ohio 44139-2794 Products: Site AnalyzerModels: SA-1700 S

Page 19

Bird Site Analyzer2Items Not Supplied! Calibration Combination! Bird 5010 Directional Power Sensor! Bird 5011 Terminating Power SensorSite Analyzer Fe

Page 20 - Measure Power Mode

Introduction3Antenna Test! Rejects on-channel interfering signals to +13 dBm.! Stores up to 12 setups.! Adjustable pass/fail limit line with visual in

Page 21 - Chapter 2 Getting Started

Bird Site Analyzer4Component Description123456789

Page 22 - SA-BATPAK

Introduction51. Soft Keys Activates the function described directly to the right of the key.2. Mode KeysMode Activates the mode menu. This menu allows

Page 23 - Power Up

Bird Site Analyzer6Right ArrowDuring:Marker Control Moves active marker right one point at a time.All other times As defined by the function descripti

Page 24

Introduction7Connection DescriptionAntennaTest PortParallelPrinter PortComputer PortRemote Power SensorDC InputCAUTION+22 dBm max. inputDo not apply R

Page 25 - Chapter 3 Calibration

Bird Site Analyzer8Display DescriptionMeasureMatchSampleDisplayM2M11.7033.00 33.18MHz1.00VSWR000, ABC, 01-JAN-2001 12:01Measure Match ModeSave/RecallA

Page 26

Introduction91. Trace Graphic display of the measurement.2. Calibration Indicator Indicates the calibration status.3. Limit Line Value Indicates the l

Page 28 - Setting the Frequency

Bird Site Analyzer10MeasurePowerSampleDisplayFwd 0.0 WattsRefl600.00Match0.00 Watts0.00 Loss(dB)Measure Power ModePwrUnitWattsSensor ConnectedMatchUni

Page 29 - Measure Match Mode

11Chapter 2 Getting StartedPower SupplyInternalBatteryThe Bird Site Analyzer has an internal, rechargeable, lithium-ion battery pack. This will operat

Page 30 - 33.00 33.18

Bird Site Analyzer12To charge the external battery pack:1. Plug the automobile cigarette lighter adapter into the external battery pack on the side th

Page 31

Getting Started13Power UpOn first power up or after a failure set the unit to default parameters. Refer to “Return to Defaults”, page 83.Self TestA se

Page 32

Bird Site Analyzer14

Page 33

15Chapter 3 CalibrationCalibration AccessoriesCalibration Combination including:! One 50 ohm load! One Open standard! One Short standardTest Cable (op

Page 34

Bird Site Analyzer16! Press the CALIBRATE key.! Attach the open standard.Press the OPEN key.Wait for a “beep” and for the trace to scroll before conti

Page 35 - Cable Loss Mode

17Chapter 4 Measure Match ModeThis measurement verifies and monitors the match conditions in the antenna system at various frequencies. The results ar

Page 36 - Limit Line

Bird Site Analyzer18Setting the FrequencyFrequencies can be set manually or chosen from a list of presets. If the start, stop, center, or span is manu

Page 37 - Limit Test

Measure Match Mode19Band ListThe band list pop-up menu contains frequency band presets. Using a preset is quick, easy, and sets test parameters while

Page 38 - Marker Adjustment

iSafety PrecautionsRemove PowerObserve general safety precautions. Do not open the instrument with the power on.Safety Earth GroundAn uninterruptible

Page 39

Bird Site Analyzer20Display theband listand select afrequencybandBand Start Stop<Full Band>LowFirePoliceEmergencyPolicePoliceEmergency25.0030.00

Page 40 - Measurement Hold

Measure Match Mode21Setting the Scale & Unit of MeasureThe display scale can be set manually or by using Auto Scale. The display can have units of

Page 41 - Printing

Bird Site Analyzer22Auto ScaleAuto Scale automatically sets the display scale so that the entire trace is displayed.Auto Scale can also be activated f

Page 42

Measure Match Mode23SettingUnitsScroll toUnitsMin: 1.00Max: 99.99Auto ScaleUnits: VSWRMeasure Match Mode25.00MHz1700.0099.991.00VSWRFreqScaleLim

Page 43 - Chapter 5 Fault Location Mode

Bird Site Analyzer24Activate theselectedunit0.0-60.0Rtn Loss (dB)25.00MHz1700.00Measure Match ModeFreqScaleLimitLineCableLossMin: -60.0 dBMax: 0.

Page 44 - Setting the Frequency Span

Measure Match Mode25Cable Loss MeasurementsTo measure cable loss, the cable being tested should have an open on the far end.Press fromMeasureMatch Mod

Page 45 - Fault Location Mode

Bird Site Analyzer26Limit LineThe limit line is an option that helps you see a failure. It appears as a horizontal line at the limit line value.Press

Page 46

Measure Match Mode27Limit TestLimit Test compares the trace to the limit line.Test Off — disables the comparison and the pass/fail indicator.Test On —

Page 47 - Setting the Cable Type

Bird Site Analyzer28Marker AdjustmentMarkers indicate VSWR to 0.01, or Return or Cable Loss to 0.1 dB. The exact values of M1, M2, and the delta marke

Page 48

Measure Match Mode29Moveacross thetraceLocate thehighestpoint of thetraceLocate thelowest pointof the traceM2M1Calibration: Full M1: 33.12 MHzM2:

Page 49

Bird Site AnalyzeriiCaution StatementsThe following equipment cautions appear in the text and are repeated here for emphasis.Safety StatementsUSAGEANY

Page 50 - Setting the Distance

Bird Site Analyzer30Measurement HoldMeasurement Hold stops tracing at the end of the current measurement and displays the last measure-ment result. It

Page 51

Measure Match Mode31PrintingThe Bird Site Analyzer can print the information displayed on the screen to any HP Deskjet printer that supports the PCL L

Page 52 - VSWR [ratio]

Bird Site Analyzer32

Page 53

33Chapter 5 Fault Location ModeThis measurement identifies the position of impedance discontinuities (faults) within the antenna/feeder system. The me

Page 54

Bird Site Analyzer34Setting the Frequency SpanThe frequency span and relative propagation velocity determine the maximum distance at which distance-to

Page 55

Fault Location Mode35For example, for a cable with a dielectric constant of 2.296, the velocity percentage is 66%. If the cable is 100 feet long, then

Page 56

Bird Site Analyzer36If you enter a span greater than the current settings allow, the span will be set to the maximum possible value. In this case, set

Page 57

Fault Location Mode37Setting the Cable TypeThe cable type can be set manually or chosen from a list of presets. If the velocity of propagation or loss

Page 58

Bird Site Analyzer38Cable ListThe cable list pop-up menu contains cable presets. Using a preset is quick, easy, and sets test parameters while elimina

Page 59

Fault Location Mode39Activate theselectedcable0.0Meters17.0Fault Location ModeVel. Prop.: 0.8200 Cable List: Loss: 0.1447 dB/m 50

Page 60

iiiUTILISATIONTOUTE UTILISATION DE CET INSTRUMENT QUI N’EST PAS EXPLICITEMENT PRÉVUE PAR LE FABRICANT PEUT ENDOMMAGER LE DISPOSITIF DE PROTECTION DE L

Page 61

Bird Site Analyzer40Setting the DistanceThe maximum possible Stop Distance is determined by the frequency span. (See Table 1 on page 34)Press fromFaul

Page 62

Fault Location Mode41SettingUnitsScroll toUnitsDisplay theunits listand selectMeters orFeet0.0Meters18.2Fault Location ModeStart: 0.0 mStop:

Page 63 - Chapter 6 Save and Recall

Bird Site Analyzer42Setting the Scale & Unit of MeasureThe display scale can be set manually or by using Auto Scale. The display can have units of

Page 64 - Save Trace

Fault Location Mode43Auto ScaleAuto Scale automatically sets the display scale so that the entire trace is displayed.Auto scale can also be activated

Page 65 - Trace Label

Bird Site Analyzer44SettingUnitsScroll toUnits0.0Meters18.2Fault Location ModeMin: 1.00Max: 99.99Auto ScaleUnits: VSWR99.991.00VSWRDistanceScaleL

Page 66

Fault Location Mode45Activate theselectedunit0.0Meters18.2Fault Location ModeDistanceScaleLimitLineSmoothLowCbl Type0.0-60.0Rtn Loss (dB) Min: -60.0

Page 67 - Save and Recall

Bird Site Analyzer46Limit LineThe limit line is an option that helps you see a failure. It appears as a horizontal line at the limit line value.Press

Page 68

Fault Location Mode47Limit TestLimit Test compares the trace to the limit line.Test Off — disables the comparison and the pass/fail indicator.Test On

Page 69

Bird Site Analyzer48Marker AdjustmentMarkers indicate VSWR to 0.01, or Return Loss to 0.1 dB. The exact values of M1, M2, and the delta marker (differ

Page 70 - Trace Label Quicktext

Fault Location Mode49Moveacross thetraceLocate thehighestpoint of thetraceLocate thelowest pointof the traceM1ActiveM2On∆MSingle0.0Meters18.2Fault Loc

Page 71

Bird Site AnalyzerivAbout This ManualThis manual covers the Bird SA-1700 and SA-1700-P Public Safety Site Analyzers.Changes to This ManualWe have made

Page 72

Bird Site Analyzer50SmoothSmooth is an option that will digitally average the displayed trace. Three levels of smooth are available.Press fromFaultLoc

Page 73

Fault Location Mode51Measurement HoldMeasurement Hold stops tracing at the end of the current measurement and displays the last measure-ment result. I

Page 74 - Trace Label Config

Bird Site Analyzer52PrintingThe Bird Site Analyzer can print the information displayed on the screen to any HP Deskjet printer that supports the PCL L

Page 75 - Save Setup

53Chapter 6 Save and RecallTraces and setups can be stored in nonvolatile memory, along with descriptive labels and a time-date stamp.Press fromMeasur

Page 76 - Recall Trace

Bird Site Analyzer54Save TracePress fromSave andRecallThe trace issavedSave and RecallSaveRecallLabelSetupDelete2 of 2:001 015 ALPHA RX02 05-OCT

Page 77 - Recall Setup

Save and Recall55Trace LabelPress fromSave andRecallThe Labelmenu isdisplayedSave and RecallDeleteMNSpaceShortcuts off. See config/help for details.OA

Page 78 - Delete Trace

Bird Site Analyzer56EnteringNumbersIf the numeric shortcuts are off, press a number key once to enter a number. If the numeric shortcuts are on, press

Page 79 - Delete Setup

Save and Recall57ChangingLabelsThe labelshould beRX01,not TX01Scroll to thetext to bechangedSave and RecallDeleteVW*Shortcuts off. See config/help for

Page 80

Bird Site Analyzer58DeletingLabelsScroll to thetext to bedeletedPress todelete thecharacter tothe left ofthe cursorSave and RecallDeleteVW*Shortcuts o

Page 81 - Chapter 7 Measure Power Mode

Save and Recall59SavingLabelsA completelabelSave and RecallDeleteVW*Shortcuts off. See config/help for details.XA D G J M P S V Y .B E H K N Q T W Z

Page 82 - Connecting a Sensor

vChapter LayoutIntroduction — Identifies the parts, functions, and features of the Site Analyzer as well as optional equipment available.Getting Start

Page 83 - Setting the Full Scale Power

Bird Site Analyzer60Trace Label QuicktextStore commonly used label elements in the Quicktext to speed label entry.ViewQuicktextIf a Quicktext entry is

Page 84 - Setting the Offset

Save and Recall61CreateQuicktextEnter textinto thelabel, thenactivateQuicktextSave and RecallSelectAdd toQtextDeleteShortcuts off. See config/help for

Page 85 - Recall Setups

Bird Site Analyzer62UseQuicktextScroll to aQuicktextentrySave and RecallSelectAdd toQtextDeleteShortcuts off. See config/help for details.ViewQTextA D

Page 86

Save and Recall63DeleteQuicktextScroll to aQuicktextentrySave and RecallSelectAdd toQtextDeleteShortcuts off. See config/help for details.ViewQTextA D

Page 87 - Setting Units

Bird Site Analyzer64Trace Label ConfigLabel HelpLabel Help provides instructions for labelling traces.LabelShortcutsEnabling shortcuts allow you to se

Page 88 - Calibrating the Bird 5011

Save and Recall65Save SetupSaving the setup will save the following:Scale Min Start Frequency Center FrequencyScale Max Stop Frequency Span FrequencyS

Page 89

Bird Site Analyzer66Recall TraceWhen a trace is recalled it is displayed along with the trace currently being measured. The frequency range will be ch

Page 90

Save and Recall67Recall Setup☞NOTE: For best results, calibrate the Site Analyzer immediately before taking measurements.Press fromSave andRecallThe S

Page 91 - Chapter 8 Utilities

Bird Site Analyzer68Delete TraceFrom Saveand Recall,scroll to thetrace to bedeletedSave and RecallSaveRecallLabelSetupDelete2 of 2:001 015 ALPHA RX0

Page 92 - Adjust Date and Time

Save and Recall69Delete SetupPress fromSave andRecallThe Setupmenu isdisplayedScroll to thesetup to bedeletedSave and RecallSaveRecallTracesDelete00

Page 93 - Return to Defaults

viTable of ContentsSafety Precautions ... iAbout This Manual ...

Page 94 - FM Modulation

Bird Site Analyzer70

Page 95 - Utility Mode

71Chapter 7 Measure Power Mode☞NOTE: Power measurement is only available with the SA-1700-P. To upgrade a standard SA-1700, please contact Customer Se

Page 96

Bird Site Analyzer72Connecting a SensorThe antenna test port is only used for testing unpowered systems. For power measurement, an external power sens

Page 97 - Chapter 9 Computer Software

Measure Power Mode73Setting the Full Scale PowerThe Bird Site Analyzer will automatically set the full scale power for sensors that do not use element

Page 98

Bird Site Analyzer74Setting the OffsetTo read unattenuated power when using a coupler or attenuator, enter (in dB) the attenuation or coupling factor.

Page 99 - Chapter 10 Maintenance

Measure Power Mode75Recall SetupsPower measurement setups can be created using the Bird Site Analyzer PCTool Software, then stored in the Site Analyze

Page 100 - Troubleshooting

Bird Site Analyzer76Choosing the Displayed MeasurementThe primary display shows one value on the large numerical display and the dial. Other measureme

Page 101 - Maintenance

Measure Power Mode77Setting UnitsTogglebetweenWatts anddBmTogglebetweenVSWR,ReturnLoss, andMatchEfficiencyFwd -30.00 dBmRefl58.00Match-30.00 dBm 2.0

Page 102 - Battery Replacement

Bird Site Analyzer78Calibrating the Bird 5011The Bird 5011 is a highly accurate terminating power sensor. Over time, the sensor’s “zero value” (readin

Page 103 - Unit Reset

Measure Power Mode79Press to thewarningFwd 0.00 mWattsOffset12.00FullScaleMeasure Power ModeConfiguration ScreenRecallSetups0.00 dB0.01 WattsCalibrate

Page 104 - Flash ROM Upgrade

viiFault Location Mode... 33Setting the Frequency Span . . . . . . . . . .

Page 105 - Parts List

Bird Site Analyzer80

Page 106 - Antenna Testing

81Chapter 8 UtilitiesThese utilities adjust the Bird Site Analyzer’s date and time, return the unit to default settings, or check the printer status.P

Page 107

Bird Site Analyzer82Adjust Date and TimePress todisplay thesystemsettingsCurrent Date / Time:12 - - 2001 12:05 pDEC Utility ModeRtn ToDefaultsFM ModP

Page 108 - Bird Site Analyzer

Utilities83Return to DefaultsAll settings will be returned to the factory presets. Saved traces and setups will not be affected. This function should

Page 109

Bird Site Analyzer84FM ModulationThe Bird Site Analyzer uses FM modulation to improve its immunity to interfering signals. Under certain circumstances

Page 110

Utilities85PrinterThe Bird Site Analyzer is compatible all printers that use HP PCL Level 3, including most HP printers.When a printer error occurs, t

Page 112

87Chapter 9 Computer SoftwareThe Bird Site Analyzer Software is designed to help you use the Site Analyzer more effectively. The software enables use

Page 113 - Limited Warranty

Bird Site Analyzer88

Page 114 - DECLARATION OF CONFORMITY

89Chapter 10 MaintenanceCleaningOnly clean the Site Analyzer with a soft cloth dampened with mild detergent and water. Do not use any other type of cl