Bird Model 6091 Calorimeter
34
InterFace Clear (IFC)
Function
Terminates all bus activity and passes control to system controller.
Remarks
All devices set to talker and listener idle states.
The National card sends IFC with the first device command.
Statements
National GPIB-PCII HP 200/300
CALL IBSIC (UD%) ABORT 7
Device Clear (DCL)
Function
Resets the status of all devices to an initialized state.
Remarks
Calorimeter returns to the default condition as listed below:
WA RF POWER MODE
YT TWO TERMINATORS (CR) (LF)
PY SEND PREFIXES
T1 TRIGGER ONE SHOT ON TALK
KY EOI ENABLED
M00 SRQ MASK OFF
Statements
National GPIB-PCII HP 200/300
MDE$=CHR$(14) CLEAR 7
CALL IBCMD (BRDO%,MDE$)